{"id":5575,"date":"2024-09-13T15:50:09","date_gmt":"2024-09-13T15:50:09","guid":{"rendered":"http:\/\/121.199.166.88\/?post_type=resources&#038;p=5575"},"modified":"2024-09-13T15:51:08","modified_gmt":"2024-09-13T15:51:08","slug":"analyzing-samples-that-occupy-less-than-10-percent-cell-volume-on-accupyc-ii","status":"publish","type":"resources","link":"https:\/\/micromeritics.com.cn\/en\/resources\/analyzing-samples-that-occupy-less-than-10-percent-cell-volume-on-accupyc-ii\/","title":{"rendered":"Analyzing Samples that Occupy Less than 10 Percent Cell Volume on AccuPyc II"},"content":{"rendered":"\n<p>Samples measured by the AccuPyc should occupy more than 10 percent of the measured cell volume for the most accurate results. This can be difficult to obtain if a low sample quantity is available or if the sample bulk density is low.<\/p>\n\n\n\n<p>The cell volume is determined through the volume calibration process and is dependent on the volumes of the AccuPyc sample chamber and sample cup. There are two methods that can be used to lower the cell volume, therefore increasing the sample occupancy percentage.<\/p>\n\n\n\n<h3 class=\"wp-block-heading\">Method 1: Using chamber inserts<\/h3>\n\n\n\n<p>Chamber inserts consist of an aluminum cylinder and a lower-volume sample cup. The aluminum cylinder occupies a region of the sample chamber which reduces the cell volume and acts as a holder for the new lower-volume sample cup. Before starting an analysis with a chamber insert, a volume calibration must be performed with the insert. Available insert sizes are listed in Table 1.<\/p>\n\n\n\n<figure class=\"wp-block-image size-full\"><img fetchpriority=\"high\" decoding=\"async\" width=\"808\" height=\"521\" src=\"https:\/\/micromeritics.com.cn\/wp-content\/uploads\/2024\/09\/table-1-nominal-chamber-sizes.png\" alt=\"AccuPyc insert sizes\" class=\"wp-image-5790\" srcset=\"https:\/\/micromeritics.com.cn\/wp-content\/uploads\/2024\/09\/table-1-nominal-chamber-sizes.png 808w, https:\/\/micromeritics.com.cn\/wp-content\/uploads\/2024\/09\/table-1-nominal-chamber-sizes-300x193.png 300w, https:\/\/micromeritics.com.cn\/wp-content\/uploads\/2024\/09\/table-1-nominal-chamber-sizes-768x495.png 768w\" sizes=\"(max-width: 808px) 100vw, 808px\" \/><figcaption class=\"wp-element-caption\">Table 1. AccuPyc insert sizes<\/figcaption><\/figure>\n\n\n\n<h3 class=\"wp-block-heading\">Method 2: Using calibration spheres<\/h3>\n\n\n\n<p>Cell volume can be reduced by using a non-porous sample with a known volume. The calibration spheres are ideal for this method since the volumes are well known and traceable. Analyses are performed with both the calibration sphere and the sample of unknown density. The following steps must be performed before starting an analysis on the AccuPyc when using this method.<\/p>\n\n\n\n<ol class=\"wp-block-list\">\n<li>Perform a volume calibration using the standard sample cup size to obtain the measured cell volume.<\/li>\n\n\n\n<li>Record the volume of the calibration sphere. If using a non-porous material that is not the calibration sphere and the volume is unknown, perform an analysis to determine the volume.<\/li>\n\n\n\n<li>Subtract the calibration spheres volume, or that of the non-porous material, from the measured cell volume.<\/li>\n\n\n\n<li>Enter the new calculated cell volume into the appropriate location. Cell volume can be found in the following locations:\n<ul class=\"wp-block-list\">\n<li>Windows software \u2013 Unit menu -> Calibration -> Calibrate Volume<\/li>\n\n\n\n<li>Keypad software \u2013 Set Up -> Calibration Data<\/li>\n<\/ul>\n<\/li>\n<\/ol>\n\n\n\n<h3 class=\"wp-block-heading\">Method Comparison<\/h3>\n\n\n\n<p>Analyses were performed on a 100 cm<sup>3<\/sup> AccuPyc using the standard sample cup, a 10 cm<sup>3<\/sup> insert for the first method, and the calibration spheres (approximately 51 cm<sup>3<\/sup>) that come with the unit for the second method. Data was collected for two separate calibration spheres, with volumes 2.421875 cm<sup>3<\/sup> and 0.718532 cm<sup>3<\/sup>. The calibration spheres used for this experiment are a non-porous tungsten carbide alloy. Table 2 shows the averages of the data collected.<\/p>\n\n\n\n<figure class=\"wp-block-image size-large\"><img decoding=\"async\" width=\"1024\" height=\"490\" src=\"https:\/\/micromeritics.com.cn\/wp-content\/uploads\/2024\/09\/table-2-measured-volumes-1024x490.png\" alt=\"Measured volumes, standard deviations, percent difference, nominal chamber size based \non method used, and percent of sample chamber occupied\" class=\"wp-image-5791\" srcset=\"https:\/\/micromeritics.com.cn\/wp-content\/uploads\/2024\/09\/table-2-measured-volumes-1024x490.png 1024w, https:\/\/micromeritics.com.cn\/wp-content\/uploads\/2024\/09\/table-2-measured-volumes-300x144.png 300w, https:\/\/micromeritics.com.cn\/wp-content\/uploads\/2024\/09\/table-2-measured-volumes-768x367.png 768w, https:\/\/micromeritics.com.cn\/wp-content\/uploads\/2024\/09\/table-2-measured-volumes.png 1392w\" sizes=\"(max-width: 1024px) 100vw, 1024px\" \/><figcaption class=\"wp-element-caption\">Table 2. Measured volumes, standard deviations, percent difference, nominal chamber size based on method used, and percent of sample chamber occupied<\/figcaption><\/figure>\n\n\n\n<p>The measured sample volume becomes more accurate as the percent of the chamber volume occupancy increases. The sample material for this experiment is non-porous. The difference between the true and measured volume should be expected to be greater when porous materials are analyzed since the bulk volume of the sample is greater than its skeletal volume due to the open porosity<\/p>\n","protected":false},"excerpt":{"rendered":"<p>Samples measured by the AccuPyc should occupy more than 10 percent of the measured cell volume for the most accurate results. This can be difficult to obtain if a low sample quantity is available or if the sample bulk density is low. The cell volume is determined through the volume calibration process and is dependent [&hellip;]<\/p>\n","protected":false},"featured_media":5789,"parent":0,"template":"","meta":{"_acf_changed":true},"methods":[26],"resource_type":[8],"class_list":["post-5575","resources","type-resources","status-publish","has-post-thumbnail","hentry","methods-density","resource_type-application-notes"],"acf":[],"_links":{"self":[{"href":"https:\/\/micromeritics.com.cn\/en\/wp-json\/wp\/v2\/resources\/5575","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/micromeritics.com.cn\/en\/wp-json\/wp\/v2\/resources"}],"about":[{"href":"https:\/\/micromeritics.com.cn\/en\/wp-json\/wp\/v2\/types\/resources"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/micromeritics.com.cn\/en\/wp-json\/wp\/v2\/media\/5789"}],"wp:attachment":[{"href":"https:\/\/micromeritics.com.cn\/en\/wp-json\/wp\/v2\/media?parent=5575"}],"wp:term":[{"taxonomy":"methods","embeddable":true,"href":"https:\/\/micromeritics.com.cn\/en\/wp-json\/wp\/v2\/methods?post=5575"},{"taxonomy":"resource_type","embeddable":true,"href":"https:\/\/micromeritics.com.cn\/en\/wp-json\/wp\/v2\/resource_type?post=5575"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}